The air in a vaccine fridge changes temperature every time the door opens. Your vaccines change much more slowly. Knowing the difference helps your team respond to real problems.
Two temperatures inside every fridge
A probe sitting in open air measures cabinet temperature. It reacts quickly: open the door and the reading rises, close it and it settles again within minutes.
Vaccines behave differently. Packaged stock holds its temperature for longer, so it warms and cools far more slowly than the air around it. A probe that simulates a vaccine tracks what your stock is actually experiencing, rather than every swing in the air.
Medisafe Plus Monitoring Unit
What the guidelines say
The National Vaccine Storage Guidelines ‘Strive for 5’ (4th edition, 2025) recognise the same principle. They recommend placing the data logger’s sensor probe in an empty vaccine box inside the refrigerator, so it measures temperature under the same conditions as your vaccines.
They also define a cold chain breach as exposure to temperatures outside +2°C to +8°C, but don’t count a single excursion of up to +12°C lasting 15 minutes or less. That allows for brief rises in air temperature, such as a door opening, that don’t put your stock out of range. The guidelines still ask you to note any such event and its cause on your temperature chart.
Why one probe isn’t enough
Air temperature only: short spikes from door openings can trigger alarms that don’t reflect a real risk to your stock. Over time, staff can start to ignore them.
Product temperature only: you may not notice a failing door seal, a blocked condenser or a struggling compressor until the problem has progressed.
Seeing both readings side by side shows your team what’s happening to the stock and to the refrigerator, so they can tell a routine door opening from a fault that needs attention.
How AQ Box handles it
AQ Box records cabinet and simulated vaccine temperature side by side, every minute. It sends SMS and email alerts to your nominated staff if readings go out of range, the power fails or the unit loses connectivity.
From any web-enabled device, your team can view current readings, minimum and maximum reports, and up to 90 days of temperature history. That makes the weekly review of logged data called for in the guidelines quick to do. Multi-site operators can see every unit across every site in one account.
AQ Box is built into every Medisafe Plus refrigerator, available as an option on AQ Medical, and can be retrofitted to refrigerators you already have.
If you see an excursion
If readings have gone outside +2°C to +8°C, follow your site’s cold chain breach procedure and the current guidelines. Based on the guidelines’ breach checklist (Appendix 3):
Cold chain breach: first steps
- Keep the vaccines in the refrigerator and make sure it's holding +2°C to +8°C.
- Label the affected vaccines "Do not use / Do not discard".
- Contact your state or territory health department as soon as possible.
- Don't discard any vaccine until your state or territory health department advises you to.
Source: National Vaccine Storage Guidelines 'Strive for 5', Appendix 3.